![Land 4500 MKII Premier Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/land/4500-mkii-premier/4500-mkii-premier_operation-and-maintenance-manual_3476555057.webp)
Model 4500 Premier Dust and Opacity Monitor
57
System Data Register Table
NOTE
Certain registers are read-only. Any attempt to write to these registers will result in the
return of an error code.
Register #
Parameter
Data
Description
Range
Write
Type
Access
25
OPLR
Float
Optical Path Length Ratio
0.1 - 1.0
NO
26
(Previously PLCF)
49
Opacity
Float
Calculated opacity
0.00 - 1.00
NO
50
51
Optical Density
Float
Calculated optical
0.00 - 3.00
NO
52
density
53
Dust Density
Float
Calculated dust
0.0 - 10 000
NO
54
density
55
Average Optical
Float
Average optical density
0.0 - 3.00
NO
56
Density
during gravimetric test
65
Cal Check OPz
Float
Cal check zero opacity
0.00 - 0.20
NO
66
value
67
Cal Check OP0
Float
Cal check opacity value
0.00 - 0.20
NO
68
69
Cal Check
Float
Cal check span opacity
0.00 - 1.00
NO
70
Upscale
value
85
Cal Timer
16-bit
Calibration timer counter
NO
(seconds)
86
Motor Timer
16-bit
Motor timer counter (5mS)
NO
87
Error Flags
32-bit
Error flags*
NO
88
104
System State
8-bit
System status*
YES
105
System Mode
8-bit
System operating mode*
NO
106
Cal Status
8-bit
Calibration status*
NO
For registers marked * see over page