Place a clear blank in the sample chamber. Press the
Q
/
ENTER
button to scan the
blank.
Insert Blank
* Continue
The screen will display
Blank Done
for about 1 second.
Blank Done
* Scan Blank
Place the reacted low standard in the sample chamber. Press
Q
/ENTER
to scan
the low standard.
Insert Lo Standard
* Continue
Place the reacted high standard in the sample chamber. Press
Q
/ENTER
to scan
the high standard.
Insert Hi Standard
* Continue
The meter will display the message “
Storing
” and return to the
EDIT USER
TEST01
menu.
Storing
EDIT USER TEST01
* New Calibration
Decimal Places
Select Units
44
SMART2 COLORIMETER 1.07