Press the
Q
/
ENTER
button to select
002 Aluminum
.
ALL TESTS
* 002 Aluminum
003 Ammonia - NLF
004 Ammonia - NLS
The SMART2 Colorimeter is ready to scan at the correct wavelength. Place the
blank in the sample chamber, close the lid and press the
Q
/
ENTER
button to scan
blank.
NOTE:
Do not keep the button depressed.
002 Aluminum
* Scan Blank
The screen will display
Blank Done
for about 1 second.
Scan Sample
will be
positioned next to
*
.
002 Aluminum
Blank Done
* Scan Blank
Place the reacted sample in the chamber, close the lid and press the
Q
/
ENTER
button to scan sample. The colorimeter will scan the sample and the results screen
will appear.
002 Aluminum
* Scan Sample
Record test result. To repeat the test, press the
Q
/ENTER
button to scan the
sample again. The last blank scanned is used to zero the colorimeter for repeated
scans. A different blank can be used by pressing the
s
button to scroll back to
Scan Blank
and then scanning another blank. Scroll with the
t
or
s
buttons
and make another selection with the
Q
/
ENTER
button. The %T or Absorbance of
the last test can be viewed by choosing
%T/Abs
. Press the
EXIT
button to escape
to previous menus.
NOTE:
The menus loop in this screen so either the
s
or
t
buttons will lead to
the menu selection needed.
002 Aluminum
0.09 ppm
* Scan Sample
Next Test
Previous Test
%/Abs
Calibrate
Scan Blank
30
SMART2 COLORIMETER 1.07