n
GENERAL TESTING PROCEDURES
The following are some step by step examples of how to run tests from the
Testing Menu
. These test procedures are designed to be used with LaMotte
SMART Reagent Systems.
LaMotte Company continuously updates the list of pre-programmed tests as the
calibrations become available. Pre-programmed calibrations can be added to
the SMART2 Colorimeter in the field. A Windows-based computer running a
Windows Operating System and an 8 pin mini-DIN/9 pin F D-submin serial
cable (order Code 1771) are required.
Call LaMotte Technical Services at 1-800-344-3100 (410-778-3100 outside
the USA) or email at [email protected] for a current list of available
calibrations and downloading instructions.
n
TESTING WITH THE LaMOTTE
PRE-PROGRAMMED TESTS
Press ON to turn on the Smart2 Colorimeter. The LaMotte logo will appear for
about 2 seconds and then the Start screen appears. Press the
Q
/ENTER
button to
start testing
.
VER 1.0
Smart2
* Start
The
MAIN MENU
will appear. Press the
Q
/ENTER
button to select
Testing
Menu
.
MAIN MENU
* Testing Menu
Editing Menu
PC Link
Press the
Q
/ENTER
button to select
All Tests
.
TESTING MENU
* All Tests
Sequence 1
Sequence 2
Press the
t
button to move to the
002 Aluminum
to
*
.
ALL TESTS
* 001 Alk - UDV
002 Aluminum
003 Ammonia - NLF
SMART2 COLORIMETER 1.07
29