A reverse font (dark background with light characters) will appear to indicate that
the reading can be adjusted. Use
s
or
t
to scroll to the concentration of the
sample, 0.10 ppm in this example.
002 Alumninum
0.09
* Calibrate
Set the calibration by pressing
Q
/
ENTER
to select
Calibrate
.
002 Aluminum
0.10
* Calibrate
Two menu choices will be offered,
Set Calibration
and
Factory Setting
.
Set the calibration by pressing
Q
/ENTER
to select
Set Calibration
; or use
s
or
t
to scroll to and select
Factory Setting
t o revert to the factory
calibration.
002 Aluminum
0.10
* Set Calibration
Factory Setting
The meter will display the message “
Storing
” and return to
002 Aluminum
test.
Storing
The calibration for
002 Aluminum
has now been standardized and can be used
for testing. The standardization can be removed by repeating the calibration and
selecting
Factory Setting
.
002 Aluminum
* Scan Sample
Next Test
Previous Test
%/Abs
Calibrate
Scan Blank
32
SMART2 COLORIMETER 1.07