CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS
The LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of
the pre-programmed tests by the user is not possible. However, a procedure to
standardize the calibration can be performed to obtain the most accurate
readings or to meet regulatory requirements.
The LaMotte Pre-Programmed tests are standardized with one standard
solution. To standardize over the full range of the test, the concentration of the
standard should be chosen from the high end of the range. Alternatively, if
samples do not cover the full range of the test, a standard should be chosen
that is close to the concentration of the samples.
The standardization procedure should be followed as often as required by
regulations and laws for compliance monitoring.
In the example below the Aluminum calibration will be standardized.
Prepare a standard solution to be tested. Use 0.10 ppm aluminum.
Use the
s
or
t
button to scroll to
002 Aluminum
. Follow instructions in the SMART2 Manual for testing the
aluminum standard. Scan the blank.
002 Aluminum
* Scan Blank
The screen will display
Blank Done
for about 1 second.
Scan Sample
will be
positioned next to
*
.
002 Aluminum
Blank Done
* Scan Sample
Place the reacted sample in the chamber, close the lid and press
Q
/ENTER
to scan
sample. The result will be displayed.
002 Aluminum
* Scan Sample
The displayed result can now be standardized. Use the
s
or
t
buttons to scroll to
Calibrate
. Press
Q
/ENTER
to select.
002 Aluminum
0.09 ppm
* Scan Sample
Next Test
Previous Test
%T/Abs
Calibrate
Scan Blank
SMART2 COLORIMETER 1.07