Insert blank, press the
Q
/ENTER
button to scan blank.
102 Abs 520
* Scan Blank
The screen will display
Blank Done
for about 1 second.
102 Abs 520
Blank Done
* Scan Blank
Insert the reacted sample. Press the
Q
/ENTER
button to scan the sample.
102 Abs 520
* Scan Sample
Record test result. To repeat the test, press the
Q
/ENTER
button to scan the
sample again. The last blank scanned is used to zero the colorimeter for repeated
scans. A different blank can be used by pressing the
s
button to scroll back to
Scan Blank
and then scanning another blank. Scroll with
t
or
s
and make
another selection with
Q
/
ENTER
. The %T or Absorbance of the last test can be
viewed by choosing
%T/Abs
. Press
EXIT
to escape to previous menus.
NOTE:
The menus loop in this screen so either
t
or
s
will lead to the menu
selection needed.
NOTE:
The Calibrate function does not work in the Absorbance mode.
102 Abs 520
0.95
* Scan Sample
Next Test
Previous Test
%T/Abs
Calibrate
Scan Blank
34
SMART2 COLORIMETER 1.07
Содержание Smart 2
Страница 1: ...SMART2 Colorimeter OPERATOR S MANUAL V 2 3 1 07 1919 MN...
Страница 2: ......
Страница 54: ...54 SMART2 COLORIMETER 1 07...