rain[e]observer
Operating Instructions
Operating Instructions
rain[e]observer Precipitation Measuring System
Syntax for measurements with aCC! or aMC!
Command
Response
aD0!
a<values><CRC><CR><LF>
a
– Sensor address
a
– Sensor address
D
– Command - Send data
<values>
– Called up data separated by
0
– Request for data from
corresponding prefix (“+” or “-”)
Buffer 0
<CRC>
– 3-digit CRC checksum
or
<CR><LF>
– End of response
1
= Buffer 1
!
– End of command
Additional measuring options
The following commands can be used to request further measurement data from the
Rain[e]
and call up with
aD0!
.
The measurement commands
aMn!
and
aMCn!
have the same format as the commands
aM!
and
aMC!
. The
same applies to the commands
aCn!
and
aCCn!
, which have the same format as the commands
aC!
and
aCC!
.
n
Function
Command
Response
1
Creation of variance data (over 4 s)
aM1!
a0031<CR><LF>
aC1!
a00301<CR><LF>
with CRC
checksum
aMC1!
a0031<CR><LF>
aCC1!
a00301<CR><LF>
2
Creation of heating data
Measurements
Unit
Buffer 0
Temperature inside
°C
Heating ON (1) / OFF (0)
Total heating capacity
%
aM2!
a0033<CR><LF>
aC2!
a00303<CR><LF>
with CRC
checksum
aMC2!
a0033<CR><LF>
aCC2!
a00303<CR><LF>
3
Creation of data: mean precipitation intensity and
max. and min. intensity of the last x minutes*.
Measurements
Unit
Buffer 0
Mean value of the last x minutes*
mm/min
Max. intensity of the last x minutes*
mm/min
Min. intensity of the last x minutes*
mm/min
The standard value for the time x is 10 minutes. It
can be changed using the
rain[e]
commander.
aM3!
a0033<CR><LF>
aC3!
a00303<CR><LF>
with CRC
checksum
aMC3!
a0033<CR><LF>
aCC3!
a00303<CR><LF>
* The recording of the data starts with transmission of the command. Call up must occur after exactly x minutes
with the command
aD0!
.
26