Lucid Series RF Signal Generator
Portable User Manual
Rev. 1.1
30
Dynamic
– Get an IP address from the DHCP server. The IP Address, Port and Subnet Mask
fields are not accessible.
x
IP Address
– Define a static IP address.
x
Port
– Define the SCPI port for communication. User should use a free port in the range 1 to 65535.
x
Subnet Mask
– Verify that the PC running Lucid software is on the same network.
x
MAC
– The device MAC address.
x
Host
– The device computer name.
x
BACK
– Click the button to return to the System dialog box.
3
3.6.5
Update
Device FW Update Sequence
1.
Prepare on a PC an USB flash memory with a folder containing two files. You can download the
files from the Tabor Electronics website at
http://www.taborelec.com/downloads
.
a.
portable.fw.tar.gz
b.
Manifest
2.
Insert the flash memory in a free USB connector on the device.
3.
Select on the device display the System tab, and then click the Update button to select the
firmware for updating the device.
Figure 3.16 System Update
4.
Wait for the operation to complete.
5.
Reboot the device manually by power-down/up.
x
BACK
– Click the button to return to the System dialog box.