Lake Shore Model 332 Temperature Controller User’s Manual
BEEP
Alarm Beeper Command
Input:
BEEP <state>[term]
Format:
n
<state>
0 = Off, 1 = On.
Remarks:
Enables or disables system beeper sound when an alarm condition is met.
BEEP?
Alarm Beeper Query
Input:
BEEP?
Returned:
<state>
[term]
Format:
n
(Refer to command for description)
BRIGT
Display Brightness Command
Input:
BRIGT <bright>[term]
Format:
n
<bright>
0 = 25%, 1 = 50%, 2 = 75%, 3 = 100%. Default = 2.
BRIGT?
Display Brightness Query
Input:
BRIGT?[term]
Returned:
<bright>[term]
Format:
n
(Refer to command for description)
CMODE
Control Loop Mode Command
Input:
CMODE <loop>, <mode>[term]
Format:
n,n
<loop>
Specifies which loop to configure: 1 or 2.
<mode>
Specifies the control mode. Valid entries:
1 = Manual PID, 2 = Zone,
3 = Open Loop, 4 = AutoTune PID, 5 = AutoTune PI, 6 = AutoTune P.
Example: CMODE
1,4[term]
– Control Loop 1 uses PID AutoTuning.
CMODE?
Control Loop Mode Query
Input:
CMODE? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<mode>[term]
Format:
n
(Refer to command for description)
CRDG?
Celsius Reading Query
Input:
CRDG? <input>[term]
Format:
a
<input>
A or B
Returned:
<temp value>[term]
Format:
±nnnnnn
Remarks:
Also see the RDGST? command.
6-28
Remote Operation