Lake Shore Model 332 Temperature Controller User’s Manual
MODE?
Remote Interface Mode Query
Input:
MODE?[term]
Returned:
<mode>[term]
Format:
n
(Refer to command for description)
MOUT
Control Loop Manual Heater Power (MHP) Output Command
Input:
MOUT <loop>, <value>[term]
Format:
n,
±
nnnnnn[term]
<loop>
Specifies loop to configure: 1 or 2.
<value>
Specifies value for manual output.
Example: MOUT
1,22.45[term]
– Control Loop 1 manual heater power output is 22.45%.
MOUT?
Control Loop Manual Heater Power (MHP) Output Query
Input:
MOUT? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<value>
Format:
±
nnnnnn[term]
(Refer to command for description)
PID
Control Loop PID Values Command
Input:
PID <loop>, <P value>, <I value>, <D value>[term]
Format:
n,±nnnnnn,±nnnnnn,±nnnnnn
<loop>
Specifies loop to configure: 1 or 2.
<P value>
The value for control loop Proportional (gain): 0.1 to 1000.
<I value>
The value for control loop Integral (reset): 0.1 to 1000.
<D value>
The value for control loop Derivative (rate): 0 to 200.
Remarks:
Setting resolution is less than 6 digits indicated.
Example: PID
1,10,50[term]
– Control Loop 1 P is 10 and I is 50.
PID?
Control Loop PID Values Query
Input:
PID? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<P value>, <I value>, <D value>[term]
Format:
±nnnnnn,±nnnnnn,±nnnnnn
(Refer to command for description)
6-36
Remote Operation