Lake Shore Model 332 Temperature Controller User’s Manual
Zone Settings (Continued)
Enter for Zone 01
Integ (I) 20.0
The Integral (I) value is entered using the numeric keypad, which includes the numbers 0
–
9, +/–, and
decimal point. Integral has a range of 0.1 to 1000 with a default of 20. Press the
Enter
key to accept the
new setting. You will see the next display.
Enter for Zone 01
Deriv (D) 0.0
The Derivative (D) value is entered using the numeric keypad, which includes the numbers 0
–
9, +/–,
and decimal point. Derivative has a range of 0 to 200 percent with a default of 0. Press the
Enter
key to
accept the new setting. You will see the next display.
Enter for Zone 01
Manual Out 0.00%
The MHP Output setting is entered using the numeric keypad, which includes the numbers 0
–
9,
+/–, and decimal point. Manual heater has a range of 0.001 to 100 percent with a default of 0. Press the
Enter
key to accept the new heater setting. Assuming the zone is controlling using Loop 1, you will see
the next display.
Select for Zone01 °®
Heater Range Off
Use the
s
or
t
key to select the Heater Range: High, Medium, Low, or Off. Press the
Enter
key to
accept the new heater range and return to the normal display. (If you are controlling using Loop 2, the
last heater range setting is omitted.) This completes the setting of Zone 01. Repeat the process for the
subsequent zones.
Operation
4-21