![Lake Shore Model 421 Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/lake-shore/model-421/model-421_user-manual_670638075.webp)
Lake Shore Model 421 Gaussmeter User’s Manual
Service
6-7
6.6 TOP OF ENCLOSURE REMOVAL AND REPLACEMENT
WARNING: To avoid potentially lethal shocks, turn off gaussmeter and disconnect it from
AC power line before performing this procedure. Only qualified personnel
should perform this procedure.
6.6.1 Removal
Procedure
1. Turn power switch Off (
O
).
2. Remove instrument power cord.
3. If attached, remove rack mounting brackets.
4. Use 5/64 hex key to remove four screws attaching top panel to unit.
5. Use 5/64 hex key to loosen two rear screws attaching bottom panel to unit.
6. Carefully remove the back bezel by sliding it straight back away from the unit.
7. Slide the top panel back and remove it from the unit.
6.6.2 Installation
Procedure
1. Slide the top panel forward in the track provided on each side of the unit.
2. Carefully replace the back bezel by sliding it straight into the unit.
3. Use 5/64 hex key to install four screws attaching top panel to unit.
4. Use 5/64 hex key to tighten two rear screws attaching bottom panel to unit.
5. If required, reattach rack mounting brackets.
6. Attach instrument power cord.
7. Turn power switch On (
l
).
6.7 EPROM
REPLACEMENT
The operating software for the Model 421 is contained on one Erasable Programmable Read Only
Memory (EPROM) Integrated Circuit (IC). The reference designator for the EPROM is U36, Lake Shore
Part Number 113-591. The EPROM has a sticker on top labeled with “M421.HEX” and the date. Use
the following procedure to replace the EPROM.
CAUTION
: The EPROM is an Electrostatic Discharge Sensitive (ESDS) device. Wear shock-proof
wrist straps (with a resistor that limits current to <5 mA) to prevent injury to service
personnel and to avoid inducing an Electrostatic Discharge (ESD) into the device. Refer to
Paragraph 6.2.
1. Follow the top of enclosure REMOVAL procedure in Paragraph 6.6.1.
2. Remove four phillips-head screws attaching transformer bracket to the Model 421 chassis.
Carefully pull transformer bracket up and sufficiently out of the way to gain access to the
Operating Software EPROM. See Figure 6-6.
3. Locate EPROM U36 (M421.HEX) on the main circuit board. Note orientation of existing IC.
M421.HEX
113-591
02.03.99
Typical EPROM
1
Match notch on
EPROM to notch
in socket
eprom.eps
4. Use IC puller to remove existing EPROM from socket.
5. Noting orientation of new EPROM, use an IC insertion tool to place new device into socket.
6. Install four phillips-head screws attaching transformer bracket to the Model 421 chassis.
7. Follow the top of enclosure INSTALLATION procedure in Paragraph 6.6.2.
Содержание Model 421
Страница 34: ...Lake Shore Model 421 Gaussmeter User s Manual 3 16 Operation This Page Intentionally Left Blank ...
Страница 54: ...Lake Shore Model 421 Gaussmeter User s Manual 4 20 Remote Operation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 421 Gaussmeter User s Manual 5 14 Accessories and Probes This Page Intentionally Left Blank ...