Lake Shore Model 421 Gaussmeter User’s Manual
Operation
3-7
3.8 RELATIVE
The relative function lets the user see small variations in larger fields. The setpoint (or center) of the
relative reading is set when pressing the
Relative
key. This captures the current reading and effectively
nulls the present field.
+1.420 kG DC
Relative On
Once the
Relative
key is pressed, the
Relative
On
message is briefly shown on the lower line of the
display. This is followed with the selected setpoint (SP) being displayed on the lower line and the
current plus or minus deviation from that setpoint on the upper line. A small delta (
°
) is also displayed
to signify the relative display.
+0.000 kG DC °
+1.420 kG SP
The relative feature also interacts with other features. When Relative and Max Hold functions are used
at the same time, the relative reading is still in the top display with proper annunciators, but the bottom
display shows the relative maximum instead of the relative setpoint.
+0.002 kG DC °
+0.010 kG MAX
Pressing
Max Hold
again turns off the maximum hold function, returning the relative setpoint to the
display. Pressing the
Relative
key turns off the relative function. The
Relative
Off
message is briefly
displayed.
3.9 ALARM
AND
RELAY
The alarm feature can be easily configured to perform several different tasks including error detection
and limit testing. The most powerful application is manual magnet testing and sorting. A display
annunciator, message display, beeper, and relay are all available to indicate an active alarm state. This
section describes all of the parameters available for the alarm feature then gives examples to illustrate
operation under different conditions.
When the alarm feature is on the alarm annunciator (
ª
) will appear in the top display. If there is an
alarm condition the annunciator will flash to indicate the alarm is active. If there is not an alarm
condition the annunciator will remain steady to indicate the alarm is in its normal state. Beeper and
relay operation follow the alarm condition in a similar way when they are in use. The alarm function is
non-latching so no action is required to reset the alarm. The alarm resets automatically when the alarm
condition is no longer present.
To turn the alarm feature on press the
Alarm
key and the following display will appear.
Select With °®
Alarm ¡On Off
Use the
s
or
t
keys to select between Alarm On and Off and press
Enter
to accept the new setting.