Lake Shore Model 421 Gaussmeter User’s Manual
3-6
Operation
3.6 ZERO
PROBE
The zero probe function is used to null (cancel) out the zero offset of the probe or small magnetic fields.
It is normally used in conjunction with the zero gauss chamber, but may also be used with an open
probe (registering the local Earth magnetic field). Users wishing to cancel large magnetic fields should
use the Relative function.
NOTE:
For best results, the instrument and probe should warm up for at least 5 minutes
before zeroing the probe, and at least 30 minutes for rated accuracy. The probe
and the zero gauss chamber should be at the same temperature.
To zero the probe in the zero gauss chamber, first allow the temperature of the probe and chamber to
be approximately equal. (A large temperature discrepancy affects the quality of the calibration.)
Carefully place the probe tip into the chamber. Orientation of the probe is not critical. Once inserted,
press the
Zero Probe
key and observe the following display.
Press Again With
Probe At Zero
Press the
Zero Probe
key. The
*CALIBRATING*
message is briefly displayed, followed by a return to
the normal display. The probe is now calibrated. For best results, periodic zeroing of the probe is
recommended.
3.7 MAX HOLD AND MAX RESET
The Max Hold function displays the largest field magnitude measured since the last Max Reset. When
the
Max Hold
key is pressed, the maximum value is shown in lower line of the display while the upper
line contains the current field reading.
+2.483 kG DC
2.544 kG MAX
The
Max Reset
key clears the hold value. The hold value is also reset upon power up or when
changing from AC or DC. Max Hold may also be used in conjunction with the Relative display
(refer to Paragraph 3.8).
In DC operation, the Max Hold feature holds the field reading that is largest in magnitude. This is
intended to monitor slowly changing signals. A field change not visible on the display can not be
recorded in DC Max Hold. The display shows only the magnitude of the largest reading.
In AC RMS operation, the maximum RMS value is held.