Lake Shore Model 421 Gaussmeter User’s Manual
Installation
2-5
2.8 INITIAL SETUP AND SYSTEM CHECKOUT PROCEDURE
The following procedure is an initial instrument setup and checkout procedure. The intent of this
procedure is to verify basic operation of the unit before beginning use for measurements.
CAUTION:
Check power source for proper voltage before connecting the line cord to the
Model 421. Also check the power setting on the window in the fuse drawer.
Damage to unit may occur if connected to improper voltage.
1. Check power source for proper voltage. The Model 421 operates with 100, 120, 220, or 240
(+5%, –10%) AC input voltage.
2. Check window in fuse drawer for proper voltage setting. If incorrect, refer to Paragraph 5.2.
3. Ensure the power switch is in the off (
O
) position.
CAUTION:
The probe must be connected to the rear of the unit before applying power to the
gaussmeter. Damage to the probe may occur if connected with power on.
4. Plug in the probe connector to PROBE INPUT. Use thumbscrews to tighten connector to unit.
5. Ensure any other rear panel connections (SERIAL I/O or ANALOG OUTPUTS) are connected
before applying power to the unit.
6. Plug line cord into receptacle.
7. Turn the power switch to the on (
l
) position. The front panel display should turn on and the following
message briefly appear.
Lake Shore 421
Field Monitor
8. The normal gaussmeter display will now appear. A typical display is illustrated below.
+ 0.001 G DC
NOTE:
For best results, the instrument and probe should warm up for at least 5 minutes before
zeroing the probe, and at least 30 minutes for rated accuracy. The probe and the zero
gauss chamber should be at the same temperature.
NOTE:
Some Lake Shore probes come with a clear plastic sleeve to protect the probe tip when not
in use. The sleeve slides up and down the probe cable. To place the probe in the zero
gauss chamber, slide the protective sleeve back, exposing the probe tip, before placing the
tip in the chamber.
9. If so equipped, slide the clear plastic sleeve back, exposing the tip of the probe. Place the probe in
the zero gauss chamber and press the front panel
Zero Probe
key. The display below appears.
Press Enter With
Probe At Zero