Lake Shore Model 421 Gaussmeter User’s Manual
Installation
2-3
2.4 LINE
INPUT
ASSEMBLY
This section covers line voltage and fuse verification in Paragraph 2.4.1, power cord in Paragraph 2.4.2,
and power switch in Paragraph 2.4.3.
2.4.1 Line Voltage and Fuse Verification
To verify proper line voltage selection look at the indicator in the window on the fuse drawer of the
line input assembly. Line voltage should be in the range shown in the specifications listed on the back
of the instrument. See Figure 2-2. If not, change the line voltage selector per instructions in
Paragraph 6.3. The fuse must be removed to verify its value, refer to the procedure in Paragraph 6.4.
Use slow-blow fuses of the value specified on back of the instrument.
2.4.2 Power
Cord
The Model 421 includes a three-conductor power cord. Line voltage is present across the outer two
conductors. The center conductor is a safety ground and connects to the instrument metal chassis.
For safety, plug the cord into a properly grounded three-pronged receptacle.
2.4.3 Power
Switch
The power switch turns the instrument On and Off and is located in the line input assembly on the
instrument rear. When
l
is raised, the instrument is On. When
O
is raised, the instrument is Off.
120
100/120/220/240 V
10% +6% Voltage
50-60 Hz 40 VA MAX
100/120V
220/240V
0.5 A T 250V
0.25 A T 250V
0.25×1.25"
5×20mm
Line Cord
Input
Power Switch
O
= Off,
l
= On
Fuse
Drawer
F-421-2-2.eps
Figure 2-2. Line Input Assembly
2.5 PROBE INPUT CONNECTION
WARNING: Some probes used with the gaussmeter have conductive parts. Never probe near
exposed live voltage. Personal injury and damage to the instrument may result.
CAUTION:
Always turn off the instrument before making any rear panel Probe Input connections.
Lake Shore probes plug into the 15 pin D-style connector on the Model 421 rear panel. Turn the
instrument off before attaching a probe. See Figure 8-3 for pin definitions.
When power is turned on, the instrument reads parameters from probe memory. The probe is ready to
use. No parameters need to be entered into the Model 421. However, the Zero Probe function should
be performed the first time a probe is used with the instrument and periodically during use.