![Lake Shore Model 421 Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/lake-shore/model-421/model-421_user-manual_670638052.webp)
Lake Shore Model 421 Gaussmeter User’s Manual
4-18
Remote Operation
Device Specific Commands (Continued)
RELS
Relative Mode Setpoint Command
Input:
RELS <setpoint>[term]
Format:
nnn.nn
<setpoint>
Relative mode setpoint value with up to 5 digits resolution.
Remarks:
New value is entered on the same field range as the old value. Setting value to zero first will
change the setting range to present display range.
RELS?
Relative Mode Setpoint Query
Input:
RELS?[term]
Returned:
<setpoint>[term]
Format:
nnn.nn
(Refer to command for description)
Remarks:
Use RELSM? to determine units multiplier.
RELSM?
Relative Mode Setpoint Multiplier Query
Input:
RELSM?[term]
Returned:
<multiplier>[term]
Format:
a
<multiplier>
µ = micro = 10
-6
, m = milli = 10
-3
, (blank) = unity, k = kilo = 10
3
Remarks:
Used with RELS? query.
SNUM?
Probe Serial Number Query
Input:
SNUM?[term]
Returned:
<serial>[term]
Format:
annnnnnnnn
<serial>
The current probe serial number format is Hnnnnn, though there is room for
up to a ten-character response.
TYPE?
Probe
Type
Query
Input:
TYPE?[term]
Returned:
<probe>[term]
Format:
n
<probe>
0 = High Sensitivity (HSE), 1 = High Stability (HST),
2 = Ultra-High Sensitivity (UHS)