Lake Shore Model 218 Temperature Monitor User’s Manual
6-24
Remote Operation
FILTER
Configure Input Filter Parameters
Input:
FILTER <input>, <off/on >, <points>, <window>
Returned:
Nothing.
Remarks: <input>
Specifies input to configure (
1-8
).
<off/on>
Specifies whether the filter function is off or on.
0
= Off,
1
= On.
<points>
Specifies how many data points the filtering function uses (
2-64
).
<window>
Specifies what percent of full scale reading limits the filtering function (
1-10
).
Reading changes greater than this percentage reset the filter.
Example: FILTER
1,
1, 10, 2[term] -
Filter input 1 data through 10 readings with 2% of full scale window.
FILTER?
Query Input Filter Parameters
Input: FILTER?
<input>
Returned: <off/on >, <points>, <window>
. Format:
n,nn,nn[term]
Remarks:
Returns input filter configuration. See
FILTER
command for returned parameter descriptions.
<input>
specifies which input to query (
1-8
).
IEEE
Configure IEEE-488 Interface Parameters
Input:
IEEE[<terminator>, <EOI enable>, <address>
Returned:
Nothing.
Remarks:
Configures parameters of the IEEE interface.
<terminator>
Specifies the terminator.
0
= <CR><LF>,
1
= <LF><CR>,
2
= <LF>,
3
= no terminator.
<EOI enable>
Disables/enables the EOI mode.
0
= Enabled,
1
= Disabled.
<address>
Specifies the IEEE address.
Example: IEEE 1, 0, 4[term] -
After receipt of the current terminator, the instrument responds to
address 4, uses <CR><LF> as the new terminator and uses EOI mode.
IEEE?
Query IEEE-488 Interface Parameters
Input: IEEE?
Returned: <terminator>, <EOI enable>, <address>
. Format:
n,n,nn[term]
Remarks:
Returns IEEE interface parameters. See IEEE command for returned parameter descriptions.
INCRV
Configure Input Curve Number
Input:
INCRV <input>, <curve number>
Returned:
Nothing
Remarks:
Specifies the curve an input uses for temperature conversion.
<input>
Specifies which input to configure (
1-8
).
<curve number>
Specifies which curve the input uses.
0
= none,
1-5
= Standard Diode
Curves,
6-9
= Standard Platinum Curves,
21-28
= User curves. Note:
Curve locations
10-20
not used.
Example: INCRV 5,6[term] -
Input
5
standard curve 6 (PT-100).
INCRV?
Query Input Curve Number
Input: INCRV?
<input>
Returned: <curve number>
. Format:
nn[term]
Remarks:
Returns the input curve number. See the
INCRV
command for parameter descriptions.
<input>
Specifies which input to query (
1-8
).
<curve number>
Specifies which curve the input uses.
0
= none,
1
-
5
= Standard Diode
Curves,
6-9 =
Standard Platinum Curves,
21-28
= User Curves. Note:
Curve locations
10-20