IEEE-488/SERIAL INTERFACE COMMANDS (ALPHABETICAL LISTING)
Q
CLS
Clear Interface Command
Input:
Q
CLS
Returned:
Nothing
Remarks:
Clears bits in the Status Byte Register and Standard Event Status Register and terminates all
pending operations. Clears the interface, but
not
the instrument. See
Q
RST
command.
Q
ESE
Configure Status Reports in the Standard Event Status Register
Input:
Q
ESE <bit weighting>
Returned:
Nothing
Remarks:
Each bit is assigned a bit weighting and represents the enable/disable status of the
corresponding event flag bit in the Standard Event Status Register. To enable an event flag
bit, send the command
Q
ESE
with the sum of the bit weighting for each desired bit. See the
Q
ESR?
command for a list of event flags.
Example:
To enable event flags 0, 3, 4, and 7, send
*ESE 143[term]
. 143 is the bit weighting sum for each bit.
Bit
Bit Weighting
Event Name
0 1 OPC
3 8 DDE
4 16
EXE
7 128
PON
143
Q
ESE?
Query the Configuration of Status Reports in the Standard Event Status Register
Input:
Q
ESE?
Returned: <ESE bit weighting>
. Format:
nnn[term]
Remarks:
The integer returned represents the sum of the bit weighting of the enable bits in the
Standard Event Status Enable Register. See the
*ESR?
command for a list of event flags.
Q
ESR?
Query Standard Event Status Register
Input:
Q
ESR?
Returned: <ESR bit weighting>
. Format:
nnn[term]
Remarks:
Queries for various Model 218 error conditions and status. The integer returned represents
the sum of the bit weighting of the event flag bits in the Standard Event Status Register.
Bit
Bit Weighting
Event Name
Bit
Bit Weighting
Event Name
0 1
OPC
4 16
EXE
2 4
QYE
5 32
CME
3 8
DDE
7 128
PON
Q
IDN?
Query
Identification
Input:
Q
IDN?
Returned: <manufacturer>, <model number>, <serial number>, <firmware date>
Format:
LSCI,MODEL218S,aaaaaa,nnnnnn[term]
Remarks:
Identifies the instrument model and software level.
Q
OPC
Operation Complete Command
Input:
Q
OPC
Returned:
Nothing
Remarks:
Generates an Operation Complete event in the Standard Event Status Register upon
completion of all pending selected device operations.