Lake Shore Model 218 Temperature Monitor User’s Manual
4-8
Operation
Curve Select (Continued)
To select a temperature response curve, press
Curve Select
. The display below appears.
Use the Data Selection keys to cycle through the inputs (1-8) for which to select a temperature
response curve. Press
Enter
when the desired input appears. The second display in the setting
sequence appears.
Use the Data Selection keys to cycle through the temperature response curves. When the desired
curve appears, press
Enter
to assign that curve to the selected input and return to the normal display.
Press
Escape
at any time to return to the normal display. The instrument retains values changed prior
to pressing
Escape
.
4.7 MATH
Simple math features are included for convenience and aid in setting up experiments. Readings can be
filtered to quiet effects of a noisy environment. Max and Min readings can be captured. A linear
equation can be applied to input data to correct system errors or improve performance of the analog
outputs. Math features can be performed on all eight sensor inputs. Each input must be configured
seperately.
4.7.1 Max/Min
The Max/Min feature simply captures and stores the highest (Max) and lowest (Min) reading taken
since the last reset. The feature will only capture from one reading source so it is important to select a
source. Max and min can be manually reset as described below. They are also reset when the
instrument is turned off or parameters related to the input are changed.
To select a source for Max/Min, press
Math
. The display below appears.
Press
Enter
to select an input for Max/Min. The second display in the Math setting sequence
appears.