Lake Shore Model 218 Temperature Monitor User’s Manual
5-4
Special Features
5.1.5 Erasing User Curves
To erase a user curve, press
Curve Entry
. The screen below appears.
Use the Data Selection keys to select
Erase Curve
, then press
Enter
. The second display in the
sequence appears.
Use the Data Selection keys to cycle to the input to which the curve applies (
1-8
). When the desired
input appears, press
Enter
. The final display in the sequence appears.
Press
Enter
to delete the specified curve and return to the normal display. Press
Escape
to cancel
the deletion and return to the normal display.
5.1.6 Viewing Standard Curves
View standard curves using the curve entry procedure. Standard curves are read-only (uneditable).
5.1.7 Copying
Curves
Temperature curves can be copied from one location inside the Model 218 to another. This is a good
way to make small changes to an existing curve. Curve copy may also be necessary if the user needs
the same curve with two different temperature limits or needs to extend the range of a standard
curve. The curve that is copied from is always preserved.
NOTE:
The copy routine allows you to overwrite an existing user curve. Please ensure
the curve number you are writing to is correct before proceeding with curve copy.
To copy a curve, press the
Curve Entry
key. Press the
s
or
t
key until you see the following
display.
Press the
Enter
key. You can press the
Escape
key anytime during this routine to return to the
normal display.