IEC 60601-1-2:2014 Table 2 Requirements:
The Aquarius Urodynamic Analyzer System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Aquarius Urodynamic Analyzer System should assure that it is used
in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
– Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV Contact
±2kV, ±4kV, ±8 kV
and ±15kV Air
±8 kV Contact
±15 kV Air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge IEC
61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage Dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0%
U
T
(100 % dip in
U
T
)
for 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
70%
U
T
(30% dip in
U
T
)
for 25 cycles
0%
U
T
(100% dip in
U
T
)
for 5 seconds
<5%
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the Aquarius Urodynamic Analyzer
System requires continued operation during
power mains interruptions, it is
recommended that the Aquarius Urodynamic
Analyzer System be powered from an
uninterruptible power supply or a battery.
Power Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Table 22: Electromagnetic Immunity