background image

 

LABORIE AQUARIUS XT/CT/LT/CTS Service Manual

 

LBL-000233 

[B] 

  43

 

Roam™ DX:  

Place the Roam™ DX on the charging bay/docking station on the cart. Make sure 
it clicks into place to ensure a proper connection (a beeping sound signifies it is 
fully connected). It will take approximately 7.5 hours to fully charge the device 
(when not in use).  
When the Roam™ DX is sampling, it will take 10.5 hours for the device to be fully 
charged.  
When the LED light on the device is green it is fully charged. 
 

NOTE: LABORIE recommends to have the device plugged in or have at least 

40% battery charge before the start of a test. 

  

Figure 33: Roam™ DX Charging 

Bay 

 

UPP: 

Plug the USB cable from the UPP stand into the front USB port under the central 
pump hub to charge the UPP.  It will take approximately 7.5 hours to charge (about 
9 hours if sampling).  
It is recommended to plug the UPP in at the end of every day to ensure a full 
charge for testing. 
 

NOTE: LABORIE recommends to have the device plugged in or have at least 

40% battery charge before the start of a test. The equipment can still be used 
during battery charge time.  

 

Figure 34: UPP 

 

 

5.7.2

 

B

ATTERY 

P

REVENTIVE 

M

AINTENANCE

 

 

Do not step on, drop, drop in water, or puncture the battery. If misuse, abuse, or damage is suspected, or any 
form of mechanical damage to the casing is visible, then discontinue use and return the battery to LABORIE. 

 

If a sudden change in the battery’s ability to hold a charge or a sudden change in battery life time is noticed, 
discontinue use and return the battery to LABORIE. 

 

It is recommended to charge all device batteries overnight to ensure proper function. Once the battery level 
drops to 40% charge, it is best to start recharging. 

 

5.8

 

COMPUTER VIRUS PROTECTION 

All computers purchased from LABORIE are virus-free before shipment and are installed with an antivirus program. It is the 
customer’s responsibility to correctly use and maintain the anti-virus program to prevent virus problems.  LABORIE is not 
responsible for any virus-related computer problems after point of delivery to the customer. 
 

5.9

 

TREATING AND DISPOSING OF PRODUCT AFTER USE 

 

After use, discard the contaminated, plastic, single-use disposables and any packaging according to your 
institution’s standard operating procedures on medical waste handling. 

 

For end of life product, waste electrical and electronic equipment should be collected separately and returned 
to the designated local recycling service. 

 

For end of battery life, disposal must be handled according to local regulations. 

 

Packaging waste should be collected separately for available national packaging collection and recycling 
services. 

Содержание Aquarius XT

Страница 1: ...LABORIE AQUARIUS XT CT LT CTS Service Manual LBL 000233 B 1 Aquarius XT CT LT CTS Service Manual...

Страница 2: ...cap and i LIST Office Reporter are trademarks of Laborie Medical Technologies Corp Windows is a registered trademark of Microsoft Corp Bluetooth is a registered trademark of Bluetooth SIG Inc VELCRO i...

Страница 3: ...id Filled Pressure Transducer Connections Priming 20 2 9 3 Catheter Setup 21 2 9 4 Fluid Filled Catheter Setup 21 2 9 5 T DOC Air Charged Catheter Setup 23 2 9 6 EMG Setup 23 3 Equipment Status and Li...

Страница 4: ...ive Maintenance 43 5 8 Computer Virus Protection 43 5 9 Treating and Disposing of Product After Use 43 5 10 Environmental Consideration of Waste Disposal 44 5 11 Preventive Maintenance Checking Calibr...

Страница 5: ...to the Transducer Cartridge 22 Figure 16 OFF to the Syringe 22 Figure 17 EMG Patch Placement 23 Figure 18 Pump Buttons and Indicators 24 Figure 19 Pump Connections 24 Figure 20 Roam DX Buttons and In...

Страница 6: ...ing the Spinning Disk Flow Transducer 29 Table 11 Calibrating Roam DX Pressure Channels and Transducers 30 Table 12 Calibrating ROAM DX EMG Channel 31 Table 13 Calibrating the Pump 33 Table 14 Calibra...

Страница 7: ...the device ABBREVIATIONS Common abbreviations used in this manual CHC Central Hub Computer CFG software configuration file 1 INTRODUCTION This manual is intended for servicing the AQUARIUS UDS system...

Страница 8: ...29 C to 60 C Humidity uncontrolled to 85 relative humidity Pump Hub P N PMP1000 Output Channels 2 Infusion 1 Bubble 1 USB type A 1 USB type B Roam DX P N RMX1000 Dimensions 5 5 14 cm L X 3 6 9 3 cm W...

Страница 9: ...Temperature 25 C 13 F to 50 C 122 F Anorectal Manometry ARM Test Capabilities Stationary Rest Squeeze Pushing Rectal Volumes Rectoanal Inhibitory Reflex RAIR Balloon Expulsion Calculations Rectal Comp...

Страница 10: ...0V COM532 EOL COMXXX Computer Laptop Windows 7 Professional 64 bit English Bilingual keyboard Computer Laptop Windows 10 Professional 64 bit AQUARIUS CT AQS2000 AQUARIUS Urodynamic Analyzer System 110...

Страница 11: ...nal 64 bit English Computer Custom PC Windows 7 Ultimate 64 bit Computer Medical Grade All In One PC 24 Widescreen Configured SOFTWARE COMPONENTS LAB1000 Laborie Software NGM Urodynamics Software Pack...

Страница 12: ...CAB150 EMG Shorting Leads Shielded TRA100 UROCAP V Uroflowmeter CHA171 Deluxe Commode Chair CHA181 Foldable Commode Chair CHA102 Accessories Funnel MSM1045 SHELF HEIGHT ADJUSTABLE POW1016 Power Cord 1...

Страница 13: ...elded w 5 pin Lemo 6ft ELE200 Electrode FSR Shielded 18 3 Lead Cables CAB150 EMG Shorting Leads Shielded MSM1064 Flow stand urocap 5 w wheels MSM1078 Flow stand urocap 5 w o wheels CHA1000 NGM uroflow...

Страница 14: ...ng components such as the pump and UPP as well as provide a reset to the digital portion of the Urodynamic processor The system is designed to automatically stop or reset when the firmware is unrespon...

Страница 15: ...e that all cables are organized and contained to prevent trip hazards Table 7 Spinning Disk Flow Transducer Setup 2 3 1 4 Flowstand For assembly instructions see the AQUARIUS Flowstand User Assembly I...

Страница 16: ...semble the Puller Stand UPP1006 prepare one 6mm and one 4mm HEX L KEY 1 Connect the Pole Assembly to the H Base Assembly using the provided Washer M8 Split Washer M8 and Screw M8 Use the 6mm HEX L Key...

Страница 17: ...rsal joint 2 so that the system can be positioned to a preferred angle On the side of the universal joint is a ball joint knob 3 which can be loosened to allow you to move and reset the nose unit in d...

Страница 18: ...place a graduated beaker on top of the transducer Figure 7 b If using a Spinning Disk Flow Transducer place the metal beaker under the flow transducer outlet Figure 8 3 Place the commode chair and fu...

Страница 19: ...pump head Position compressible portion of pump tubing across rollers from left to right 5 Close pump head 6 Open the thumbwheel on the pump tubing 2 8 ROAM DX SETUP FOR PRESSURE AND EMG Roam DX Conn...

Страница 20: ...IONS PRIMING Do not plug in or unplug any transducers when a procedure is running Always stop the procedure before plugging in or unplugging transducers Follow the instructions provided to prime the t...

Страница 21: ...men catheter to record bladder pressures 1 Connect the infusion line to the bladder filling port of the catheter 2 Connect the P1 measurement tubing to the vesical port of the catheter Use a triple lu...

Страница 22: ...8 Depress the Syringe plundger to infuse water back into the catheter Remove the syringe 9 Connect the primed measurement tubing to the balloon catheter To prime a double lumen catheter so it is free...

Страница 23: ...and position urethral sensor at the desired point in the urethra often at the highest pressure area 10 Connect the vesical catheter to the pump tubing NOTE Ensure bladder and rectal pressures are wit...

Страница 24: ...onnected blinking green connected and sampling orange System fault blinking orange not connected light off system off 3 Stop button Press to stop pump Press and hold for 5 seconds to reset Figure 18 P...

Страница 25: ...decrease selected value or scroll through options Start Stop OK 2 Manual wake up from hibernation Press for 5 seconds to enter changing settings mode Single push Exit from changing settings mode Pres...

Страница 26: ...ery voltage 4 2 full 3 6 empty and charger status Battery Full Charging or Battery Powered no external power System Information Unit Roam serial number Ver Firmware version Table 8 Functions of Roam B...

Страница 27: ...PP Press and hold for 5 seconds or longer to bring to sleep 3 Battery Status Indicator green battery full and plugged into charger blinking green battery good orange charger plugged and battery chargi...

Страница 28: ...edures Run the calibration procedure in one continuous process to ensure proper function 4 1 HOW TO CALIBRATE FLOW AND VOLUME 4 1 1 CALIBRATE FLOW AND VOLUME WITH THE UROCAP V 1 Tap the Calibration ic...

Страница 29: ...sducer device 3 The Calibration will open for the Spinning Disk Flow Transducer Follow the on screen instructions to complete the calibration NOTE The Next button in the final screen does not need to...

Страница 30: ...Catheter System set up for calibration 4 The calibration window for the selected Catheter System will load in this example Water based Disposable Catheter System has been selected Tap the boxes correl...

Страница 31: ...icon on the home screen 2 Tap to select the HUB device 1 Tap the Calibration icon on the home screen 2 Select the Roam DX device 3 Tap to select Electromyography Calibration The calibration window for...

Страница 32: ...calibration name For example Saline8f This name is required to setup the pump for a test 5 Tap Next to continue 6 Select the pump roller from the combo box and tap Next 7 Select a Urocap V device from...

Страница 33: ...d 11 Tap OK to stop the pump and tap Next to continue 12 Follow the on screen instructions to complete calibration NOTE During calibration there will be 8 calibration steps 13 Once complete the pump c...

Страница 34: ...on icon on the home screen 2 Tap to select the HUB device 3 Tap to select the Infusion Transducer Calibration option 4 Select a transducer and tap Next 5 Type in the Low Load value and then tap Next 6...

Страница 35: ...HE UPP PULLER NOTE Before starting UPP calibration gather a ruler and pencil 1 Tap the Calibration icon on the home screen 2 Tap to select the UPP device 3 Follow the on screen instructions to complet...

Страница 36: ...device module via USB but fails The device s USB cable may not be plugged in Communication Error This occurs when the Central Hub Computer CHC successfully connects via Bluetooth or USB but fails to c...

Страница 37: ...a broken connection and or low battery then make sure that the connection is re established or the battery is fully charged For information on LED lights refer to the Equipment Status and Light Signa...

Страница 38: ...its and odor Clean the device as follows Quick Cleaning Daily Slowly pour approximately 1L of clean water into the funnel Keep the disc running to flush the transducer cup Thorough Cleaning Weekly 1 D...

Страница 39: ...s into any liquid Keep them dry at all times Do not use solvent or abrasive cleaning agents Make sure that liquids do not enter the motor through apertures in the enclosure Do not use Chlorine Bleach...

Страница 40: ...op button to stop the motor 3 Slide the mobile clamp down towards the catheter guide Figure 27 Figure 27 Mobile clamp moving toward catheter guide 4 Once the mobile clamp is flush to the catheter guid...

Страница 41: ...Corporation 1 800 548 4873 o ENDOZYME The Ruhof Corp 1 800 537 8463 Follow the manufacturer s instructions Reference to any specific brand name is not an endorsement of their efficacy 5 5 2 2 Remove...

Страница 42: ...isposable components The following is recommended after each patient use Air charged catheters are single patient use only Discard after use Replace the tethered protective cap to the transducer housi...

Страница 43: ...e abuse or damage is suspected or any form of mechanical damage to the casing is visible then discontinue use and return the battery to LABORIE If a sudden change in the battery s ability to hold a ch...

Страница 44: ...s required Once calibration is complete perform a re check by following steps 1 to 5 listed here Contact LABORIE service for assistance if problems continue after the calibration and re check Check Fl...

Страница 45: ...and down to your side You will notice minimal EMG activity relaxed 4 Keep your palm open and lift your arm slightly at a 45 degree angle and hold the position to keep it at a constant rate You will n...

Страница 46: ...cmH2O threshold then calibration is required Once calibration is complete perform a re check by following the instructions provided above Contact LABORIE service for assistance if problems continue af...

Страница 47: ...ice for assistance if problems continue after the calibration and re check 5 12 SAFELY MOVING THE CART 1 Before moving the cart always make sure that all system components are disconnected from the ma...

Страница 48: ...on No power on electrical outlet Damaged power cord Power cord not connected properly Ensure electrical outlet is working power cords are not damaged and all power cords are secure at both ends USB p...

Страница 49: ...sables Flush lines and reset zeroes Make sure all stopcocks are in the correct position Refer to the Consumables and Transducer Setup on page 20 PUMP Pump not running Pump is not Calibrated Calibrate...

Страница 50: ...reen reads Calibration Failure Volume delta out of range Less than 500mL of water poured into beaker 500mL of water must be poured into the beaker 6 1 1 BLUETOOTH CONNECTION Unable to connect via Blue...

Страница 51: ...re i LIST MsgQListener is running To Check MsgQListener 8 Close all open programs 9 Make sure i LIST Message Queue Listener is running by clicking Start Programs Laborie i LIST MsgQListener 10 Restart...

Страница 52: ...21 Plug the power cable back into the power brick 22 Press the power button on the computer to switch the PC on again PRINTER Printer LED does not light up Printer power cable unplugged Plug in the pr...

Страница 53: ...eps 23 Go to C UDS120 and open the HK folder 24 Make sure the file format icons are set to the AutoHotKey application as shown to the right 25 If the icons are not set to the AutoHotKey application fo...

Страница 54: ...seconds to reset the system Table 19 Troubleshooting NOTE The Laborie Aquarius LT CT CTS and XT are designed to be used as stand alone cart systems with the only mains power supply connection going th...

Страница 55: ...l scientific and medical ISM radio frequency equipment Electromagnetic disturbance characteristics Limits and methods of measurement CISPR 22 2008 Information technology equipment Radio disturbance ch...

Страница 56: ...equipment RF Emissions CISPR 11 Class A The Aquarius Urodynamic Analyzer System is suitable for use in all establishments other than domestic and may be used in domestic establishments and those direc...

Страница 57: ...kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply inpu...

Страница 58: ...nsmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of known RF transmitting devices and...

Страница 59: ...LABORIE AQUARIUS XT CT LT CTS Service Manual LBL 000233 B 59 Table 23 Electromagnetic Immunity Radio Frequency Communications...

Страница 60: ...ed no closer than 30 cm 12 inches to any part of the Aquarius system including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Rated maximu...

Страница 61: ...icates the humidity range to which the medical device can be safely exposed Temperature Limit 5 3 7 1 Indicates the temperature limit to which the medical device can be safely exposed Atmospheric Pres...

Страница 62: ...ams component part lists and other technical documentation directly related to the AQUARIUS UDS system 10 END USER SOFTWARE LICENSE AGREEMENT Find licenses and terms for LABORIE products and services...

Отзывы: