LABORIE AQUARIUS XT/CT/LT/CTS Service Manual
LBL-000233
[B]
56
IEC 60601-1-2:2014 Table 1 Requirements
The Aquarius Urodynamic Analyzer System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Aquarius Urodynamic Analyzer
System should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 1
The Aquarius Urodynamic Analyzer System uses RF energy only for
its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The Aquarius Urodynamic Analyzer System is suitable for use in all
establishments other than domestic, and may be used in domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures, such as re- orienting
or relocating the Aquarius Urodynamic Analyzer System or shielding
the location.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions IEC
61000-3-3
Complies
Table 21: Electronic Emissions