LABORIE AQUARIUS XT/CT/LT/CTS Service Manual
LBL-000233
[B]
34
4.4
HOW TO CALIBRATE THE INFUSION TRANSDUCER
1.
Tap the Calibration icon on the home screen:
.
2.
Tap to select the ‘HUB’
device:
.
3.
Tap to select the Infusion
Transducer Calibration
option:
.
4.
Select a transducer and tap Next.
5.
Type in the
Low Load value and then tap Next.
6.
Continue by following the on-screen instructions until calibration
is complete:
.
Table 14: Calibrating the Infusion Transducer