LABORIE AQUARIUS XT/CT/LT/CTS Service Manual
LBL-000233
[B]
53
Symptom(s)
Possible Cause(s)
Check/Corrective Action(s)
CONTROL PANEL
Script buttons
(Uroflow Summary,
Summary Review,
etc…) stopped
working?
Program file for formatting
buttons is not set
properly?
The program used for
formatting script buttons must
use the AutoHotKey.ahk
format. To verify the program
is correct follow these steps:
23.
Go to C:\UDS120 and
open the HK folder.
24.
Make sure the file
format icons are set to
the AutoHotKey
application as shown
to the right:
25.
If the icons are not set
to the AutoHotKey
application, for
example they are in
the Notepad format
(as seen to the right)
then you must change
it to the AutoHotKey
application:
26.
Right-click on any line
and select Properties.
In the
Opens with
section click the
Change button and
select the
AutoHotKey Unicode
application in the
resulting window and
then click OK:
.
.
.
27.
The
Opens with section will display the AutoHotKey
application.
28.
Click OK.