Axius Part No: 6137120-795
Issue: 1.0
Printed On: June 10, 2023
Axius
Installation
(continued)
CAUTION
Loose screws
Damage and injury
The tightening torque for all screws of size M8 or larger is 10 Nm! Screws situated on the inside can
be tightened using an Allen key or an open-ended spanner (SW 6 mm).
CAUTION
Overload
Damage and injury
The maximal load of the support system must not be exceeded! The following values apply to a
column height of up to 3.9 m:
Max. Load
Vertical
Horizontal
Total
1200 N
1000 Nm
support arm
300 N
240 Nm
Mounting plate
300 N
350 Nm
Dual arm
600 N
500 Nm
Bracket
300
N
240
Nm
Note
The lower fastenings must be installed first.
Hook-in the support arm(s)/mounting plate(s) at the top of the column profile and position
and tighten them at the desired height as requested by the customer.
User spirit level to align the support arms during the assembly on the column such as to be
vertical.
The following heights (upper edge of fastening) and relative positions of the various Labomed
assemblies are recommended where these are to be combined with a dental unit from Labomed:
Assembly for
Height
Position
Axius
2400 – 3900 mm
9 o’clock
Monitor arm adapter
1490 mm
9 0’clock
Tray Adapter
1230 mm
9 o’clock
X-Ray adapter
1675 mm
9 o’clock
Microscope
2512
mm
12
o’clock
Operating light
1900 mm
9 0’clock
Per definition, these dimensions relate to the distance between floor and upper edge of the support
arm on the column. The four media channels of column
⑤
, which each are situated between two T-
slots, are intended for cables or other media lines.