Axius Part No: 6137120-795
Issue: 1.0
Printed On: June 10, 2023
Axius
Installation
(continued)
CAUTION
Improper installation
Injury and damage
After the installation, check the support system for secure footing and ability to bear load.
5.6 Floor plans
Using the Paper template provided,
mark the positions of the three
mounting holes on wooden
membrane.
Drill three 1/4”(6mm) pilot holes
either direct into Joist or the
additional wooden membrane as per
the roof available.
Tighten all the bolts using Allen key
10mm.
C
Room without
inserted ceiling
Using the Paper template provided,
mark the positions of the three
mounting holes on wooden membrane.
Drill three 1/4”(6mm) pilot holes
either direct into Joist or the additional
wooden membrane as per the roof
available.
Tighten all the bolts using Allen key
10mm.
Floor Perfora
Ɵ
on
Contour Cap
Cutout contour of the inserted ceiling
External contour cap
Room With Inserted ceiling
Installa
Ɵ
on plan and drill template
Drill the three mounting holes 14mm
diameter and 100mm deep.
Place a 10mm flat washer and a split washer
over each of the three M10 X 75 hex head
RAWL Plug.
Tighten all the bolts with 10mm allen key.
External Contour cap
View from below of the room
ceiling (without inserted ceiling)
Inserted ceiling with cutout
24