Mars-Mars PRO
User Manual
Code No. 9.900.900.710
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Mars Mars Pro manual-UK - rev. 06.doc
Page 29 af 34
Remove the rectangular steel frame placed on the cabinet’s ceiling. Now the exhaust filter is
accessible; lift it up and place it in a plastic bag for biohazard waste. Insert the new filter by
performing the reverse operations for its installation.
Filters must be provided with gasket on both sides.
Before installing the new filters, the filter frame must be cleaned very carefully.
The new filters must correspond to make and the model indicated on the tags placed on the removed
panel (also see paragraph 11.2 LIST OF SPARE PARTS).
A ROUTINE TEST IS COMPULSORY AFTER THE REPLACEMENT OF THE FILTERS
(see dedicated paragraph).
The prefilters on Mars Pro are easily exchanged while the cabinets are running under full speed to
ensure full operator and environment protection. The prefilter cassettes are taken up and the prefilter
are carefully placed in plastic back and closed safely. Before installing the new filters, the filter frame
must be cleaned very carefully. Be careful not to harm the prefilters.
The new filters must correspond to make and the model indicated on the tags placed on the removed
panel
12.2 Part list for Filter
The cabinet has been tested and approved with main and exhaust filters from Camfil.
Pre filter (Mars Pro): Camfill Megalam MD-G10-VH14
Main filter: Camfil Megalam MG14-G10-1220×054×110-01PUH14
Exhaust filter: Camfil Megalam MD5014-33-G10-610×054×069-01PUH14
12.3 Fuses
The fuses of Mars GS are located in the right side of the top of the cabinet behind a protective cover.
Component
Fuse
Phase/zero
Main
T-10A
Ø5x20
Main
T-10A
Ø5x20
Electrical outlet
T-6.3A
Ø5x20
Electrical outlet
T-6.3A
Ø5x20
Fan 1
T-5A
T-5A
Ø5x20
Ø5x20
Fan 2
T-5A
Ø5x20