Mars-Mars PRO
User Manual
Code No. 9.900.900.710
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Mars Mars Pro manual-UK - rev. 06.doc
Page 15 af 34
7.3 Connections
Required fusing: Circuit breaker 10 A or fuse T 10 A.
In addition, the applicable safety requirements of the local power Supply Company shall apply.
If an automatic relay for disconnection of power in case of electric fault is needed,
it must be installed in the supply, as it is not built-in.
Before connecting the power supply it must be checked that the mains specifications
correspond to those stated on the type plate.
The safety cabinet is provided with a flexible power cord. The connection may be
installed hard-wired or by means of a wall outlet with protective ground.
If a hard-wired connection is used, a main switch, which will cut off all poles to the unit, must be
used. This switch must be lockable both in the ON and OFF positions.
If the unit is connected through a protective-ground wall outlet, the connector has the all-pole
insulator function. The wall outlet is to be installed out of reach of operators and may only be
accessible to authorized personnel.