Procedure:
With a spinning disc the test aerosol (Potassium iodide 1.5% in Ethanol) is spread inside the Mars-
Mars Pro, in a homogeneous aerosol cloud. The particles leave the disc in a horizontal course, and try
to pass the air curtain.
Four Air Samplers each with a filter membrane, catch particles which escape through the air curtain.
With a developing agent (Palladium chloride 0.1%) particles deposited on the filter membrane will
show up as brown spots. The amount of spots is to be counted.
Acceptance:
A
pf
=1×10
5
– Refer to EN 12469 for details.
9. Work Rules
9.1 Before start-up
The cabinets may only be operated at temperatures between 15
0
C and 35
0
C, at maximum 80%
relative humidity, and at normal air pressure.
All cabinets are developed and produced for use in clean environments.
They must not be operated outdoors or in environments with extreme air pollution. The safety
cabinets are not intended to filter vapours containing acids or organic solvents. The safety
cabinets must not be used as a fume hood.
Approximately 15 minutes before any work in the cabinet, the fan of the unit must be switched on
at normal velocity.
The work chamber and the front- and side windows are to be carefully cleaned and disinfected.
Use an ethanol solution or similar. It is recommended to use special lint-free material. Do not use
explosive disinfectants. See also section 10.
Objects and appliances must be carefully cleaned or disinfected before being introduced into the
work chamber. Do not bring in writing utensils, packing material, etc.
The front window is positioned in working position and kept in that position during the entire
work process.
Necessary appliances for use during work must be placed within easy reach.
Secure the appropriate protection of the operator as well as the product (e.g. clothes, gloves, etc.).