6-18
Echo Liquid Handler User Manual
6.2.7 Replace the AC power fuse
The Echo main AC power entry module includes two-user replaceable
fuses.
Use only the correct, specified fuse type as a replacement. The Echo AC
power configuration and replacement main AC power fuse are specified on
the unit’s product label on the rear panel.
Table 6.3. AC power configuration and fuses
1.
Turn off the AC power switch, if it is not already off.
2.
Unplug the AC cord from the Echo power entry module.
3.
Pry open the power entry module fuse cover with a small screwdriver.
4.
Extract the fuse holder with a small screwdriver.
5.
Replace the two AC power fuses. See
Figure 6.9 Replacing the AC power fuses
6.
Return the fuse holder to the power entry module.
7.
Close the fuse cover.
8.
Plug in the power cable.
9.
Turn on the Echo liquid handler.
Echo AC power configuration
Replacement AC power fuse
115 VAC, 60 Hz, 10 A
115 VAC, 10 A, FAST (
P/N
001-3966)
230 VAC, 50 Hz, 5 A
250 VAC, 5 A, FAST (
P/N
001-3967)
Содержание Echo 520
Страница 1: ...For Echo models 520 550 555 ...
Страница 20: ...xx Echo Liquid Handler User Manual ...
Страница 58: ...2 14 Echo Liquid Handler User Manual ...
Страница 82: ...3 24 Echo Liquid Handler User Manual ...
Страница 100: ...4 18 Echo Liquid Handler User Manual ...
Страница 160: ...5 60 Echo Liquid Handler User Manual ...
Страница 192: ...A 8 Echo Liquid Handler User Manual ...
Страница 199: ......