5-40
Echo Liquid Handler User Manual
5.
Enter the wells to be surveyed (default value is the entire plate). Use one
of the following methods:
Enter the values in the
Parameters
section.
a. Enter a valid starting well. Example:
A1
.
b. Enter the number of rows to be surveyed. Example:
16
c. Enter the number of columns to be surveyed. Example:
24
Use the cursor to select the wells in the plate map.
6.
Set up the following options in the
P
ARAMETERS
section as appropriate:
Save Results to Disk
. Select this checkbox to automatically store the
results of the acoustic survey in the client PC data directory (C:\Program
Files\Labcyte\Medman\Data). Also see File | Save on page 5-42.
Ignore Source Plate Sensor
. Select this checkbox to resolve problems
with fitting a source plate onto the source plate gripper stage. This
option allows the Echo liquid handler to use a source plate without
detecting its presence with the source plate sensor.
Figure 5.36 Diagnostic survey plate map
7.
Click the
G
O
button. The acoustic survey will be performed. The acoustic
transducer moves in an efficient, non-stop path beneath each specified
well. Acoustic pulses are transmitted up into a well, and the resulting
return data is saved for analysis.
Содержание Echo 520
Страница 1: ...For Echo models 520 550 555 ...
Страница 20: ...xx Echo Liquid Handler User Manual ...
Страница 58: ...2 14 Echo Liquid Handler User Manual ...
Страница 82: ...3 24 Echo Liquid Handler User Manual ...
Страница 100: ...4 18 Echo Liquid Handler User Manual ...
Страница 160: ...5 60 Echo Liquid Handler User Manual ...
Страница 192: ...A 8 Echo Liquid Handler User Manual ...
Страница 199: ......