Chapter 3: Getting Started
Product Service 1-800-522-7658
17
Original instructions
The identification plate, model number, serial number, and electrical connection
boxes are accessible from the front of the fume hood by removing the front panel.
The Protector Radioisotope Hood is normally wired for 115 Volt, 50/60 Hz, 20
Amp or 230Volt, 50/60 Hz, 10 Amp electrical service. Check the I.D. plate behind
the front panel for voltage verification. The number of circuits varies depending on
the model. All of the electrical connections are terminated at the field wiring
terminal box for hook-up by a qualified electrician. We recommend each circuit
be a dedicated branch circuit. However, if wired together the maximum load
allowed is the sum of individual outlets plus the rating of the unit (i.e. 2 Amps
remote, 8 Amps integral blower) not to exceed the 10A or 20A electrical service.
The single point internal junction box is used for the connection of the lights,
blower, and duplex outlets. Refer to the wiring diagram for your Protector
Radioisotope in
Appendix C: Protector Radioisotope
Specifications.
The fume hood is required to be grounded to the MAINS protective earthing ground
for safe operation. Using a ring terminal sized for a 10-24 machine screw, connect the
MAINS ground conductor to the grounding lug marked with the protective earthing
symbol,
. Only MAINS ground conductors should be connected to the protective
earthing ground lug, no other conductors should be connected to this grounding lug.
Using wire nuts connect the MAINS supply conductors to the fume hood supply
wires. Insure that the wires are connected as per the appropriate wire color codes for
the input voltage. For 115V Phase (Hot) is black and Neutral is white, for 230V
Phase1 is brown and Phase2 is blue. Refer to the wiring diagram for your Protector
Radioisotope in
Appendix C: Protector Radioisotope
Specifications.
Содержание Protector 1204 Series
Страница 50: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 46 Original instructions Figure B 2...
Страница 51: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 47 Original instructions Figure B 3...
Страница 52: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 48 Original instructions Figure B 4...
Страница 54: ...Appendix C Protector Radioisotope Specifications Product Service 1 800 522 7658 50 Original instructions...
Страница 55: ...Appendix C Protector Radioisotope Specifications Product Service 1 800 522 7658 51 Original instructions...