Chapter 8: Troubleshooting
Product Service 1-800-522-7658
41
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Vertical-rising sash
frame is distorted.
Place horizontal sliding glass
symmetrically and pull sash down to air
foil. Straighten damaged frame.
Cable is frayed or has
slipped off the
pulleys.
Re-install, cable must be replaced
immediately if damaged.
Electrical duplex
outlets no longer
have power
Wires not connected
or faulty duplex.
Check wire connection or replace duplex.
Circuit breakers
tripped in building
electrical supply.
Reset circuit breakers.
Service valves no
longer operate
Faulty building
supply.
Inspect building supply shut off valves
and appropriate pressures below the
recommended 40 PSI.
Valve no longer
operates.
Replace valve and check for leaks.
Supply line or outlet
line has leaks.
Inspect line for leaks and fix any leaking
plumbing connections.
Spray Nozzles no
longer work
Poor water supply
Inspect building water supply.
Clogged nozzle
Replace spray nozzles.
Содержание Protector 1204 Series
Страница 50: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 46 Original instructions Figure B 2...
Страница 51: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 47 Original instructions Figure B 3...
Страница 52: ...Appendix B Protector Radioisotope Dimensions Product Service 1 800 522 7658 48 Original instructions Figure B 4...
Страница 54: ...Appendix C Protector Radioisotope Specifications Product Service 1 800 522 7658 50 Original instructions...
Страница 55: ...Appendix C Protector Radioisotope Specifications Product Service 1 800 522 7658 51 Original instructions...