98
Step Skipped
SS
Segment Skipped by
operator.
No action
required.
Program will drain
and begin next
programed
segment.
Temp Sensor
TS
Temperature Sensor
failure.
Check Sensor
Screen for temp
reading.
If no reading and
wiring connected
replace sensor.
Open buttons or
display on control
panel inoperable or
controls
malfunction
Microprocessor
memory error.
Perform "Factory
Reset."
Keypad or display
defective.
Replace key pad or
display or IOB.
Detergent cup in
door fails to open
Glassware
interfering with cup
door.
Move glassware.
Malfunction of cup.
Detergent Cup
Test: with door
closed
Fail: If cup is getting
power and wiring
has continuity
replace Detergent
cup.
D-cup latch fouled.
Manually test door
operation.
Clean D-cup latch
Poor drying
Dry temperature too
low.
Program warmer dry
temperature.
Flask, beaker, or
labware with
concave bottoms not
loaded correctly.
Tilt these articles
when loading so that
water drains off.
Recondensation.
If labware is left in
the washer,
recondensation may
occur. Open latch
or program Auto
Unlock after
completion of cycle
to allow moisture-
laden air to escape.
Dry time too short.
Program longer
time.
Heater(s)
disconnected or
malfunctioned
Heater/s Test:
Reconnect or
replace heater.
Blower disconnected
or malfunctioned.
Dry Bower Test.
Reconnect or
replace blower.