94
10: Troubleshooting
This section details common troubleshooting for your Glassware Washer.
Refer to the following table if your Glassware Washer fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco Product
Service for additional assistance.
Problem
Alarm Code Possible Cause
Diagnostic
Test
Possible
Corrective
Action
Low Auto Rinse Aid
A1
Rinse Aid level low.
Check Rinse Aid
level.
Refill or replace
Rinse Aid bottle.
Float movement
restricted.
Check Sensor
Screen for reading
and float function.
Clean float on dip
tube.
Bad wiring
connection.
Check wire
connections.
Correct wiring.
Auto Rinse Aid not
used.
Turn Auto Rinse Aid
dispense off in
Accessories.
Dismiss alarm.
Conductivity
Sensor
CS
Conductivity sensor
disconnected.
Check wire
connections.
Correct wiring.
Bad conductivity
sensor.
Check Sensor
Screen for a
conductivity value.
If no reading and
wiring connected
replace conductivity
sensor.
Low Auto Detergent
D1
Auto Detergent level
low.
Check Auto
Detergent level.
Refill or replace
Detergent bottle.
Float movement
restricted.
Check Sensor
Screen for reading
and float function.
Clean float on dip
tube.
Bad wiring
connection.
Check wire
connections.
Correct wiring.
Auto Detergent not
used.
Turn Auto Detergent
dispense off in
Accessories.
Dismiss alarm,
Drain Fail
DF
Clogged Sump
screen.
Visual inspection. Clean sump screen.
Содержание FlaskScrubber 4110010 Series
Страница 74: ...74 Figure 7 1 B DETAIL B SLIDE BRACKET RACK HANDLE UPPER RACK LOWER RACK RACK BRACKET PIN SEATS...
Страница 101: ...101...