36
Current Reading (uncorrected)
– This gives the raw sensor reading with no
offset applied to the reading.
Offset
– This is the value that will be added to the Current Reading
(uncorrected) to give the proper adjusted reading that matches the reference
sensor being used for calibration. The adjusted reading (uncor offset)
is the value that will be presented on all screens outside of the Sensor
Calibration screen. Pressing the Offset field of the desired sensor will bring up
a dialog box for entering the offset value for the selected sensor.
CALIBRATION
Offset:
Cancel
Calibrate
Calibrate
– Pressing this button will save the entered offset value.
Temperature sensor offset values can range from -
5 to +5⁰C. Vacuum sensor
offset values can range from -0.020 to +0.020 mbar.
Display
This screen contains settings for controlling various functions of the touch screen
display.