![Labconco 700821500 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/labconco/700821500/700821500_user-manual_3167969031.webp)
31
NTP server
– Enter the “web address” (URL) of the NTP server of your choice
that will provide clock synchronization data to the freeze dryer via ethernet
connection.
Daylight Savings (ON/OFF)
– This feature can be used to adjust the time for
areas observing Daylight Savings Time (DST).
Units
Sele
ct your desired unit of measurement for Temperature (⁰C or ⁰F) and Vacuum (mbar,
Pa, or Torr).
Security
A four-digit security code can be set that will require users to enter the code to gain
access to selected screens. A security code will always be required to enter the
“Security” screen.
The factory default security code is “1234”
System
– If the System security feature is selected (box checked), users will
be required to enter the security code to gain access to the touch screen
controls when the screen is reactivated from a screen timeout (sleep mode as
described in Settings/Display section) or power failure.
Programs, Email Alerts, Calibration, Network
– Users will be required to
enter the security code to gain access to each individual screen that has been
selected.
Alerts
If a valid network connection has been established (via the Settings/Network screen)
and Email Settings have been properly configured, alerts can be emailed to selected
Содержание 700821500