![Labconco 700821500 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/labconco/700821500/700821500_user-manual_3167969028.webp)
28
New Program
When the “New” button is selected, the following screen will be displayed:
Pressing the program name field will allow you to edit the program name. The step
numbers are listed in numerical order. Shelf Temp., Time, and Vacuum can be edited
by pressing any of these cells. The “Add” button will add a new step to the program.
Pressi
ng the “remove” button will delete the selected step from the program. Changes
will not be saved until the “Save” button is pressed. Pressing the “Back” button will
return to the Programs screen without saving changes.
Shelf Temp.
– May be set anywhere from -20⁰C to +60⁰C in 1⁰C increments.
Time
– This time represents how long the system will hold at the Shelf Temp.
that has been programmed for each step. Time can be set from 00:01 to 99:59 (1
min to 99 hrs 59 min). It can also be set to an indefinite amount of time, which is
represented by the infinity symbol (∞).
Vacuum
– Vacuum control can be set from 0.000 to 1.500 mbar.