HM-JACKarc Operation Manual
Functions of Each Screen
4 - 2 Rev.4
Analysis
icon
Touch this icon to go to Order screen to start preparations for
measurement. See 4.1.3 Order for details.
Cell Set
icon
Touch this icon to move to the Cell Set screen to display the status for
Cells on the Reaction Table. See 4.1.5
“Cell Set” for details.
Wash
icon
Touch this icon to run a wash or prime the system.
After the initialization of the Injection Arm, the system performs priming to
fill the tube between the Sample Pipette and the Sample Syringe with
Detergent, and the tube between the Reagent Pipette and the Reagent
Syringe with the Buffer. Then the Analyzer stops.
When the Buffer or Detergent have been replaced or replenished, touch
this
Wash
icon to run a wash/prime cycle.
Home position
icon
Touch this icon to return all moving parts to their home positions.
Calibration curve
icon
Touch this icon to go to Calibration screen. Master Curve/Calibrator
registration are available from the screen.
See 4.1.5
“Calibration” for details.
Results
icon
Touch this icon to display analysis results. The Results screen can be
used to display or print results, or send them to the host computer. To
display real time results, touch this icon during operation.
See 4.1.7
“Result” for details.
QC
icon
Touch this icon to display analysis results for quality control.
X-R control diagrams and histograms are available from the menu.
See 4.1.8 QC for details.
Emergency STOP
icon
Touch this icon to stop all movement immediately.
Using this function
wastes all the processing measurement.
For actions after an emergency stop, refer to page 5-28
“Actions for
Recovery from Emergency STOP
”.
Applied Calibration curve Number
The currently applied Calibration curve Number (No. 1 in this case) is
displayed.