HM-JACKarc Operation Manual
Analysis Operations
5 - 28 Rev.4
2.
Touch
Master
button at the bottom of the Calibration screen.
3.
Touch a column of I.S.T for one of the Points from 1 to 7 and touch
ENTER
.
4.
A numeric keypad appears. Input corresponding value referring to the
Master Curve Card. Repeat process 3 and 4 until all the Points are input.
5.
After input, confirm that the input data are correct according to the Master
Curve Card.
6.
After confirmation, touch
Clear
column in the User Parameter screen to
change it back to
Protect
to avoid any unexpected change.
■
Inputting Expiration information of Master Curve
1.
After clearing input protection, touch Calibration on the Main Menu.
2.
Touch
Master
button at the bottom of the Calibration screen.
3.
Touch
Exp.Edit
button at the bottom of the Calibration Master Edit screen.
4.
A numeric keypad appears. Input Expiration information (MM.YY) referring
to the Master Curve Card and touch
ENTER
.
5.
After input, confirm that the input data are correct according to the Master
Curve Card.
6.
After confirmation, touch
Clear
column in the User Parameter screen to
change it back to
Protect
to avoid any unexpected change.
■
Inputting Lot information of Master Curve
1.
After clearing input protection, touch Calibration on the Main Menu.
2.
Touch
Master
button at the bottom of the Calibration screen.
3.
Touch
Lot
button at the bottom of the Calibration Master Edit screen.
Column of I.S.T for Point 1
Column of I.S.T for Point 7