HM-JACKarc Operation Manual
Analysis Operations
5 - 18 Rev.4
Preparation of Cup Measurement Rack (option)
This rack is used to measure samples from Sample Cups. Set the
Adaptors (white blocks) on the Rack and place Sample Cups with
samples onto the Adaptors.
Replicate of individual Cup can be configured from 1 to 3 in Item
Parameter.
For manual mode, CUP needs to be chosen in attribution and the
position has to be selected to set the orders.
There is no special instruction required to Auto measurement mode but
Analyzer recognizes the Adaptor for presence of Cups. Remove the
unnecessary Adaptors prior to measurement.
Handling Precautions: If bubbles remain at the bottom (the cap side) or midway in the Sample
Collection Picker, accurate measurement values might not be obtained.
:
The Adaptors should be placed with Cups with samples.
Rack barcode
Orientation of
sample barcode
Sample rack
Samples set on the Rack
Note
NG
(measured without
samples)
OK
(cups
measured)
OK
(not
measured)
Rack with Adaptors and Cups