10.
Troubleshooting
WARNING
The interventions mentioned may be dangerous, and must therefore only be carried out by suitably
trained technicians.
The most frequent problems are indicated below. If a problem arises that is not listed in the table,
contact the Technical Support Service.
Problems
Cause
Solutions
When the button is
activated, the lift
boom/patient
stander does not go
up/down.
Generic
-
Check that there is nothing blocking the mobile
parts of the patient lift.
-
Check that the electrical wiring is connected
properly.
-
Check the electrical wiring for damage or
interruptions.
-
Check the state of charge of the battery pack. If
the LED turns red when the keys of the
-
push‐button control panel are pressed, this
indicates that the battery pack
-
has no charge. Charge the battery.
-
Check if the lift boom is folded.
-
If the boom is not folded, replace the electric
motor or the battery pack.
The actuation and
release lever of the
patient lift legs does
not go back to the
locked position.
Generic
-
Check that there is nothing hindering the
actuation lever of the patient lift.
-
Check that the linkages are correctly secured
and that the spring
-
is sufficiently compressed.
-
If necessary, remove the plastic plug and use a
flat 19 mm spanner to compress the spring and
tighten the nut. Refit the plastic plug.
In the hydraulic
version, the boom
does not move
when the lifting
lever is operated.
No oil in the oil‐
hydraulic actuator
Descent valve in
the open position.
-
Check there are no oil leaks
-
Close the valve
If, despite all the checks made and steps taken to resolve the problem, the device
63
continues to have anomalies that limit (even partially) correct operation, the Operator and
person assigned to device use must contact the authorised Technical Support Service, the
dealer or the Manufacturer without delay.
62
10.
Troubleshooting
WARNING
The interventions mentioned may be dangerous, and must therefore only be carried out by suitably
trained technicians.
The most frequent problems are indicated below. If a problem arises that is not listed in the table,
contact the Technical Support Service.
Problems
Cause
Solutions
When the button is
activated, the lift
boom/patient
stander does not go
up/down.
Generic
-
Check that there is nothing blocking the mobile
parts of the patient lift.
-
Check that the electrical wiring is connected
properly.
-
Check the electrical wiring for damage or
interruptions.
-
Check the state of charge of the battery pack. If
the LED turns red when the keys of the
-
push‐button control panel are pressed, this
indicates that the battery pack
-
has no charge. Charge the battery.
-
Check if the lift boom is folded.
-
If the boom is not folded, replace the electric
motor or the battery pack.
The actuation and
release lever of the
patient lift legs does
not go back to the
locked position.
Generic
-
Check that there is nothing hindering the
actuation lever of the patient lift.
-
Check that the linkages are correctly secured
and that the spring
-
is sufficiently compressed.
-
If necessary, remove the plastic plug and use a
flat 19 mm spanner to compress the spring and
tighten the nut. Refit the plastic plug.
In the hydraulic
version, the boom
does not move
when the lifting
lever is operated.
No oil in the oil‐
hydraulic actuator
Descent valve in
the open position.
-
Check there are no oil leaks
-
Close the valve
If, despite all the checks made and steps taken to resolve the problem, the device
62
10.
Troubleshooting
WARNING
The interventions mentioned may be dangerous, and must therefore only be carried out by suitably
trained technicians.
The most frequent problems are indicated below. If a problem arises that is not listed in the table,
contact the Technical Support Service.
Problems
Cause
Solutions
When the button is
activated, the lift
boom/patient
stander does not go
up/down.
Generic
-
Check that there is nothing blocking the mobile
parts of the patient lift.
-
Check that the electrical wiring is connected
properly.
-
Check the electrical wiring for damage or
interruptions.
-
Check the state of charge of the battery pack. If
the LED turns red when the keys of the
-
push‐button control panel are pressed, this
indicates that the battery pack
-
has no charge. Charge the battery.
-
Check if the lift boom is folded.
-
If the boom is not folded, replace the electric
motor or the battery pack.
The actuation and
release lever of the
patient lift legs does
not go back to the
locked position.
Generic
-
Check that there is nothing hindering the
actuation lever of the patient lift.
-
Check that the linkages are correctly secured
and that the spring
-
is sufficiently compressed.
-
If necessary, remove the plastic plug and use a
flat 19 mm spanner to compress the spring and
tighten the nut. Refit the plastic plug.
In the hydraulic
version, the boom
does not move
when the lifting
lever is operated.
No oil in the oil‐
hydraulic actuator
Descent valve in
the open position.
-
Check there are no oil leaks
-
Close the valve
If, despite all the checks made and steps taken to resolve the problem, the device