3
COMMISSIONING / OPERATION
28
VFC 200
www.krohne.com
06/2019 - 4005059401 - AD VFC200 PB-PA R01 en
3.3.2 Mapping of DIAGNOSIS_EXTENSION bits into DIAGNOSIS bits if "Classic Status and
Diagnosis" handling selected
DIAGNOSIS_EXTENSION (Condensed)
DIAGNOSIS (Classic)
0
1
2 3
0
5
2
5
7
Octet number
Bit n
u
mber
Subparameter
Hardware f
ailure of the elec
tronic
Failure in measurement
Co
nfig
uratio
n no
t v
alid
M
aint
enace requir
ed
M
ore diagnosis available
0
0
Event group 24: C - Process
X
X
1
Event group 25: C - Configuration
X
X
2
Event group 26: C - Electronics
X
X
3
Event group 27: C - Sensor
X
X
4
Event group 28: F - Process
X
X
5
Event group 29: F - Configuration
X
X
6
Event group 30: F - Electronics
X
X
7
Event group 31: F - Sensor
X
X
1
0
Event group 16: M - Process
X
X
1
Event group 17: M - Configuration
X
X
2
Event group 18: M - Electronics
X
X
3
Event group 19: M - Sensor
X
X
4
Event group 20: S - Process
X
X
5
Event group 21: S - Configuration
X
X
6
Event group 22: S - Electronics
X
X
7
Event group 23: S - Sensor
X
X
2
0
Event Group 8: Config. - Process
1
Event Group 9: Config. - Initial
2
Event Group 10: Config. - High Flow
(F)
(C)
(S) / (M)
X
3
Event Group 11: Config. - Low Flow
(F)
(C)
(S) / (M)
X
4
Event Group 12: Config. - Power Failure
(F)
(C)
(S) / (M)
X
5
Event Group 13: Config. - Operation Info
6
Event Group 14: reserved
7
Event Group 15: reserved
Table 3-17: Mapping of DIAGNOSIS_EXTENSION bits into DIAGNOSIS bits if "Classic Status and Diagnosis" handling selected