PROFIBUS SETTINGS
4
37
VFC 200
www.krohne.com
06/2019 - 4005059401 - AD VFC200 PB-PA R01 en
4.4 Menu "D Service"
This menu is protected. You will need to use the service password to gain access.
4.5 Status messages and diagnostic information
Function
Settings / descriptions
D4.0 Service Parameter
D4.0 Service Parameter
D4.0 Service Parameter
D4.0 Service Parameter
D4.7.0 Identification No.
Same as C5.8.6
D4.8.0 PB Cold Start
Initiates a PROFIBUS cold start.
During a PROFIBUS cold start nearly all parameter values of the device will
be set to their default values except for the PROFIBUS address and number.
The PROFIBUS cold start will be carried out without disconnecting from a
PROFIBUS master system.
Select:
No (terminates this function) / Yes (PROFIBUS cold start will be carried out
at once)
Table 4-4: Description of menu "D Service"
NAMUR groups
(fixed level)
NAMUR groups
(configurable
level)
Group
no.
Event
no.
Event name
Description
F_Electronics
30
17
FB Electronic Failure
The fieldbus main board has an
electronic failure.
F_Electronics
30
16
FB Invalid Assembly/SN The fieldbus assembly is not
correct.
F_Electronics
30
15
FB (PA) Runtime Data
Failure
Runtime data failure occurred
in FB Main.
F_Electronics
30
14
FB (PA) Runtime
execution failure
Runtime execution failure
occurred in FB Main.
F_Configuration
29
24
FB Parameter update
failure
Failure occurred during
reading of parameters from
the Support MCU.
F_Configuration
29
23
FB NV data failure
Data failure detected in NVRAM
of FB Main which cannot be
repaired.
F_Configuration
29
22
FB Configuration failure Configuration failure in FB
Main.
F_Configuration
29
21
FB Passivated
Fieldbus application is
passivated.
F_Process
28
31
FB DeviceVariable
BadProcessPressure
The status of the fieldbus
device variable "Pressure" is
set to "BAD - process related".
F_Process
28
30
FB DeviceVariable
BadProcess
Temperature
The status of the fieldbus
device variable "Temperature"
is set to "BAD - process
related".
C_Configuration
25
29
FB Simulation Active
Fieldbus Sim. Active
C_Configuration
25
28
FB DeviceVariable
BadFunctionCheck
Pressure
The status of the fieldbus
device variable "Pressure" is
set to "Bad - function check".
C_Configuration
25
27
FB DeviceVariable
BadFunctionCheck
Temperature
The status of the fieldbus
device variable "Temperature"
is set to "Bad - function check".
S_Electronics
22
30
FB Module out of spec.
FB Main out of spec.