GG-11008 Edition 1.2
- 55 -
6-4.1 Serial number
Enter and check the serial No.
Standards
Test procedure
Tools
The serial No. shall be
correct (five least significant
digits)
1. Press Serial number and enter the serial number.
2. Check the serial No. displayed next to Serial
number to see if it is correct.
3. Press OK to return to the jig software menu
screen and press Exit to exit the jig software.
4. Press OFF to terminate the software.
5. Restart the jig software.
6. Open the system menu screen and check that the
correct serial No. appears next to Serial number.
Visual
check
6-4.2 Configured data preservation
Save the current settings to the flash memory of the panel PC. (They are not saved to the flash
ROM of the main board.) Only one type of settings can be saved. In the second and subsequent
saving operations, the data is overwritten.
6-4.3 Configured data reading
Read the settings from the panel PC flash memory.
6-4.4 Configured data clear
Clear all settings (readjustment). When readjusting, save data (see Section 3.3.3.2 "Saving
Data") in advance. If readjustment has failed, data can be restored (see Section 3.3.3.3 "Reading
Data").