Error code
Display message
Phenomenon / Solution
CA0A001
Failed to export .
This System is busy .
Please output after rebooting the system .
CA0A002
The device is not ready .
Please remove and insert the device again .
CA0A003
The specified device cannot be writen.
Please confirm the lock/unlock status of the device. If
locked, please unlock the device and reinsert it .
CA0A004
There is not enough free space on the device .
Please make more free space and then reinsert the device .
CA0A005
Failed to delete . Please try again .
Please try again . If an error has occurred again, please try
again after rebooting the system .
CA10001
A command with the same name already exists .
A command with the same name has been registered . Do
the registration again .
CA10002
Intonation file settings failed.
Failed to set intonation file. Do the registration again.
CA10003
Can't add any more commands.
The number of registered commands has reached the up-
per limit . Delete other commands and then do the registra-
tion again .
CA16001
DO NOT use the Biopsy during the battery-pow-
ered operation .
Please connect the AC adapter and then use the Biopsy .
CA1F001
System may exceed the MI/TI/Ispta .3 described
in the user’s manual . Transmission voltage was
lowered .
The system can be used as is . If you have the error again,
please contact your service representative .
CA1F002
System error has occurred .
please contact your service representative .
Please contact your service representative .
CA2B001
The system has failed to load User Account Set-
tings files. The system has been unstable.
Are you sure you want to restore to the default
setting?
Please select [Yes] and set User Account again .
CA60001
An internal error has occurred .
If an error has occurred again after rebooting,
please contact your service representative .
Please reboot the system .
If an error has occurred again after rebooting, please con-
tact your service representative .
CA60002
Database connection was failed .
So that system is unstable, please reboot the
system .
CA60003
External storage device was failed to be ejected .
Please disconnect the external storage device
after shutting down the system .
Please disconnect the external storage device after shutting
down the system .
CA61001
The required tag or data is not found .: [tag name]
Please check the connected device .
Please contact your service representative .
CA61002
Internal parameter error has occurred .
Please contact your service representative .
Please contact your service representative .
CA62003
Store Error .
Please start a new study to store images . If a store error
occurs again, try again after rebooting the system .
Содержание SONIMAGE HS2
Страница 1: ...EN 01...
Страница 2: ......
Страница 6: ...6...
Страница 7: ...7 Introduction...
Страница 41: ...41 For all product inquiries contact your service representative Support...
Страница 42: ...42...
Страница 43: ...43 Chapter 1 Product Overview...
Страница 68: ...68...
Страница 69: ...69 Chapter 2 Before Examination...
Страница 84: ...84...
Страница 85: ...85 Chapter 3 Starting Examination...
Страница 122: ...122...
Страница 123: ...123 Chapter 4 Diagnosis Mode...
Страница 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Страница 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Страница 232: ...232...
Страница 233: ...233 Chapter 7 Maintenance Inspection...
Страница 243: ...243 Chapter 8 Transducer and Biopsy...
Страница 270: ...270...
Страница 271: ...271 Chapter 9 MI TI...
Страница 283: ...283 Chapter 10 Acoustic Output...
Страница 286: ...286...
Страница 287: ...287 Chapter 11 Others...
Страница 305: ...305...
Страница 306: ...306...
Страница 307: ......