265
Chapter 8
Attaching the optional Acoustic Standoff to a trans-
ducer enables acoustic coupling by creating an ap-
propriate gap while still allowing close contact between
the patient's body and the transducer, and is helpful in
adjusting focus when examining small parts and mus-
culoskeletal structures near the surface of the body .
(The standoff amount is approx. 5 mm.)
• Do not use it for non-diagnostic purposes .
• Do not allow any person other than the dedicated
staff specified by the medical facilities to use it.
• Maintain / manage it on your own responsibility .
•
Acoustic Standoff expires thirty (30) days after it is
opened . Discard Acoustic Standoff if it is past the
expiry date .
• Write the date when you opened the package on the
label on the storage bag .
• The warranty period is thirty (30) days from the date
when the storage bag is first opened. We will replace
it at no cost if any abnormality considered to be
clearly attributable to us, occurs during this period .
• Use it in a safe and proper manner, and take full ad-
vantage of the function .
8 .3 .1 Safety Precautions &
Warnings for Acoustic
Standoff
Before using, thoroughly read the instruction manual,
and observe the following safety precautions and
warnings .
Typical risks are shown in this document . You should
be careful about other risks as well .
DANGER
•••••••••••••••••••••••••••••••••••••
• Do not use Acoustic Standoff in a location where
flammable gases or high concentrations of oxygen
gas occur. Otherwise, it may catch fire, explode or
burn .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
8 .3
Acoustic Standoff
WARNING
•••••••••••••••••••••••••••••••••••••
• Follow the instructions below . There is a risk of in-
fection .
– Keep it from coming into direct contact with any
body part (organ, mucous membrane, blood and
body fluid) other than body surfaces.
– When cleaning or disinfecting it, avoid using
unspecified cleaning / disinfecting agents.
– Make sure to clean / disinfect Acoustic Standoff
before use . Wear gloves before cleaning /
disinfecting it . Before attaching it to the transducer,
wear disinfected / sterilized gloves .
– Always clean it before disinfecting it .
– After cleaning / disinfecting Acoustic Standoff,
thoroughly rinse its surface with sterile water .
–
When you have used Acoustic Standoff, be sure to
clean / disinfect it before re-using .
– If there is a possibility that Acoustic Standoff
contacts an organ or mucous membrane or
is soiled with blood or body fluids, attach the
transducer covers . Replace it with a new one if it is
soiled with blood or body fluids.
• When Acoustic Standoff is attached to the trans-
ducer and the system is on standby, make sure to
freeze the image . Keeping the image in a live state
may cause the transducer to generate heat, result-
ing in deterioration or burns of Acoustic Standoff.
• Do not drop or give a shock to the transducer with
Acoustic Standoff. Otherwise, an electric shock or
damage may occur .
• Do not use it for surgical procedures .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
CAUTION
•••••••••••••••••••••••••••••••••••••
•
Avoid attaching Acoustic Standoff to the transducer
above the patient . It can cause injury if dropped .
• It is not sterilization-resistant . Do not sterilize it .
Otherwise, Acoustic Standoff may deteriorate.
• It is not plasma processing-ready . Acoustic Stand-
off, if subjected to plasma processing, may deterio
-
rate .
•
Make sure to observe the specified storage condi
-
tions . Otherwise, the performance can degrade and
no longer be covered under the 30-day warranty .
•
To store Acoustic Standoff, use the protective film
and bag that were originally supplied with it . If it is
stored in a manner other than that specified herein,
it may deform or discolor contact objects .
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Содержание SONIMAGE HS2
Страница 1: ...EN 01...
Страница 2: ......
Страница 6: ...6...
Страница 7: ...7 Introduction...
Страница 41: ...41 For all product inquiries contact your service representative Support...
Страница 42: ...42...
Страница 43: ...43 Chapter 1 Product Overview...
Страница 68: ...68...
Страница 69: ...69 Chapter 2 Before Examination...
Страница 84: ...84...
Страница 85: ...85 Chapter 3 Starting Examination...
Страница 122: ...122...
Страница 123: ...123 Chapter 4 Diagnosis Mode...
Страница 181: ...181 Chapter 5 Common Function in Ultrasound Images...
Страница 219: ...219 Chapter 6 Reviewing Printing and Transferring Images...
Страница 232: ...232...
Страница 233: ...233 Chapter 7 Maintenance Inspection...
Страница 243: ...243 Chapter 8 Transducer and Biopsy...
Страница 270: ...270...
Страница 271: ...271 Chapter 9 MI TI...
Страница 283: ...283 Chapter 10 Acoustic Output...
Страница 286: ...286...
Страница 287: ...287 Chapter 11 Others...
Страница 305: ...305...
Страница 306: ...306...
Страница 307: ......